Case study illustrates undetected focal atrial tachycardia in an 81-year-old male, indicating device limitations.

Key Points

  • Focal atrial tachycardia went undetected by the dual chamber ICD, leading to missed diagnosis.
  • ICD interrogation did not identify the sustained atrial tachycardia due to programmed detection thresholds.
  • Electrophysiologist successfully terminated the tachycardia through pacing after clinical evaluation.
  • Patient required ablation therapy following recurrence, emphasizing the need for device monitoring.

Cite This Study

Jagadish et al. (2025) studied this question.

synapsesocial.com/papers/68e7d631bd66d359be62663ahttps://doi.org/10.36347/sjmcr.2025.v13i10.007
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Syncope due to Pacing Inhibition During VT Detection in a Conduction System Pacing ICD Programmed in AAI(R) Mode2026
  2. 2Incessant Ventricular Fibrillation in a Patient with a Left Ventricular Assist Device and an Implantable Cardioverter-Defibrillator: A Case Report2026
  3. 3Predictors of Atrial High‐Rate Episodes and Inappropriate Device Therapy in Implantable Cardioverter Defibrillators2025
  4. 4S-ICD Oversensing of Leadless Atrial Pacing Spikes2026
  5. 5Head-to-Head Comparison of Single- Versus Dual-Chamber ICD Discriminators for Tachyarrhythmia Detection: A Single-Manufacturer, Remote Monitoring-Based Bicentric Study2025