Ancient Principles, Modern Impact: Increased Femoral Offset Demonstrates Decreased Load to Failure in Total Hip Arthroplasty

View Full Paper
Ask AI
Bookmark
Share

Authors

WSWilliam F. ShermanJTJackson P TateNRNolan M Reinhart

Discussion

Loading...

Member takes

Overview

Experimental analysis shows that femoral offset significantly influences load to failure in implants, indicating potential fracture risks.

Key Points

  • Higher femoral stem offset reduces load to failure, increasing the risk of fractures.
  • Implants with a 132° neck angle showed an average load to failure of 3,538 N compared to 2,948 N for a 127° neck angle.
  • Biaxial testing was applied to 20 osteoporotic femur models to assess mechanical viability under load.
  • Findings suggest a need for caution in using high offset stems despite benefits in hip stability.

Cite This Study

Sherman et al. (2025) studied this question.

synapsesocial.com/papers/68d43edd713b0b5dfea7e780https://doi.org/10.5435/jaaosglobal-d-25-00255
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Femoral Offset and Hip Function in Total Hip Prosthesis2025
  2. 2Femoral and Global Femoral Offset, but not Anteroposterior Offset, to Improve Postoperative Outcomes Following Total Hip Arthroplasty: Considerations Independent of the Contralateral Side.2025
  3. 3Assessing Implant Stability in Cementless Femoral Components With Different Interference Fits2025
  4. 4Periprosthetic Femur Fractures in Hemiarthroplasty Correlated With Stem Type: An Analysis From the American Joint Replacement Registry2025
  5. 5The Femoral Neck‐Shaft Offset: A Key Element in the Reconstruction of Intertrochanteric Fracture2025