Towards Single-Shot Whole-Heart T1 Mapping Within Four Seconds

View Full Paper
Ask AI
Bookmark
Share

Authors

DMDaniel MacknerMBMoritz BlumenthalNSNick Scholand

Discussion

Loading...

Member takes

Overview

Innovative method demonstrates accuracy of T1 mapping in whole heart, suggesting enhanced diagnostic effectiveness.

Key Points

  • High-resolution multi-slice T1 quantification can be completed in just four seconds, reducing patient discomfort and motion artifacts.
  • Phantom studies showed inaccuracies in T1 mapping measurements were limited to an average difference of less than 8 ms.
  • Using radial single-shot IR-FLASH technique, the innovative scanning approach was validated through phantom and in-vivo evaluations.
  • The method enhances clinical feasibility of whole-heart T1 mapping by allowing for effective imaging during shorter breath-holds.

Cite This Study

Mackner et al. (2025) studied this question.

synapsesocial.com/papers/68d42934713b0b5dfea6ea2chttps://doi.org/10.58530/2025/0813
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Accelerated myocardial T1 mapping at low field2025
  2. 2A fast 2D free-breathing myocardial T1 mapping technique with multi-slice free-running acquisition and self-navigated motion calibration2025
  3. 3Free-breathing High-resolution 3D Whole-heart Adiabatic T1ρ Mapping at 0.55T2025
  4. 4Rapid 3D Whole-Brain High-Resolution T1 Quantification: Accelerating Standard Inversion Recovery with Stack-of-Spirals FLASH Acquisition2025
  5. 5Free-breathing rapid T1 mapping for the dynamic response quantification of contrast-enhanced whole-heart myocardium2025