September 10, 2025Jentashapir Journal of Cellular and Molecular BiologyOpen Access

Impact of Lipemia and Hemoglobin S Variant on HbA1c Measurement: A Comparison of Diazyme Enzymatic Method and Capillary Electrophoresis

View Full Paper
Ask AI
Bookmark
Share

Authors

ZLZahra LotfiSFSomayeh FarahmandSZSedigheh Zargani

Discussion

Loading...

Member takes

Overview

Analytical study compares HbA1c accuracy in diabetes using enzymatic method and capillary electrophoresis under lipemia, suggesting increased reliability of CE.

Key Points

  • Capillary electrophoresis proves to be more reliable than the enzymatic method for measuring HbA1c under severe lipemia.
  • Significant false elevation in HbA1c was observed with the enzymatic method at high lipemia (P < 0.01).
  • Saline washing did not mitigate lipemic interference in the enzymatic method, leading to further inaccuracies.
  • Capillary electrophoresis maintained stable results, indicating its higher resistance to matrix interferences.

Cite This Study

Lotfi et al. (2025) studied this question.

synapsesocial.com/papers/68c23caeb210217d6478a8f9https://doi.org/10.5812/jjcmb-162424